Determination of Cadmium by Graphite Furnace Atomic Absorption Spectrometry Using Ammonium Sulphate-Tris modifier
-
-
Abstract
A method has been developed for the determination of Cd in enviromental and biological samples by platform graphite furnace atomic absorption spectrometry. Ammonium sulphate and tris-(hydroxymethyl)-aminomethane (Tris) were employed as a chemical modifier. The tolerable pyrolysis temperature for Cd was raised to 850℃ in the presence of modifier, the sensitivity for the determination of Cd was also improved if compared with that obtained by using (NH4)3PO4-Mg(NO3)2 and (NH4)2PO4-HNO3 as chemical modifier.
-
-